ASTM D5453 Biodiesel Analysis Equipment Ultraviolet Fluorescence Sulfur Analyzer

Product Details

Place of Origin: CHINA

Brand Name: SKYLINE

Certification: CE , Calibration Certificate

Payment & Shipping Terms

Minimum Order Quantity: 1 Unit

Price: negotiated

Packaging Details: plywood case

Payment Terms: L/C, D/P, T/T, Western Union, MoneyGram

Supply Ability: 10 unit / month

Get Best Price
Product Details
Highlight:

oil analysis equipment

,

oil analysis machine

Name:
Ultraviolet Fluorescence Sulfurbiodiesel For Biodiesel
Standard:
ASTM D5453
Solid Sample:
1 ~ 5mg
Liquid Sample:
1 ~ 20uL
Measuring Range:
0.5 ~ 10000mg
Power Requirements:
AC220V ± 22V, 50Hz ± 0.5Hz, 1500W
Brand:
SKYLINE
Keywords:
Biodiesel Analysis Equipment
Name:
Ultraviolet Fluorescence Sulfurbiodiesel For Biodiesel
Standard:
ASTM D5453
Solid Sample:
1 ~ 5mg
Liquid Sample:
1 ~ 20uL
Measuring Range:
0.5 ~ 10000mg
Power Requirements:
AC220V ± 22V, 50Hz ± 0.5Hz, 1500W
Brand:
SKYLINE
Keywords:
Biodiesel Analysis Equipment
Product Description

Biodiesel Analysis Equipment Ultraviolet Fluorescence Sulfur Analyzer

ASTM D5453 Biodiesel Analysis Equipment Ultraviolet Fluorescence Sulfur Analyzer 0

Application Of Biodiesel Analysis Equipment

Fluorescence sulfur analyzer is the most advanced sulfur element analyzer. The total sulfur content is determined by ultraviolet fluorescence method, and the main part is imported, which makes the performance of the whole machine comparable to similar imported products. Instruments are widely used in petroleum and petrochemical, import and export commodity inspection, bio-pharmaceutical, environmental protection, health, teaching, scientific research and other departments.

 

Product description

The instrument is suitable for the determination of total sulfur content of paraffin oil, diesel, kerosene, gasoline, lubricating oil, fuel oil, liquefied gas and natural gas, as well as other oil products, chemical raw materials and finished products.

 

Standards

SH/T 0689-2000 Determination of total sulfur content of light hydrocarbons and engine fuels and other oils (Ultraviolet fluorescence method)

ASTM D5453 The standard test method for the determination of the total sulfur content of light hydrocarbon, spark ignition engine fuel, diesel engine oil and engine oil.

GB/T 11141-2014 Determination of trace sulfur in industrial light olefins

 

Performance characteristics

1. The main part of the instrument adopts imported detector to make the detection limit comparable to imported products.

2. The localization of instrument sample oxidative cracking, etc., the daily maintenance costs are greatly reduced.

3. Imported purification device to effectively isolate carbon deposits.

4. The instrument adopts the most advanced micro-detection technology (sample gas constant temperature heating technology) to ensure the repeatability and accuracy of small samples.

5. The instrument has extremely high sensitivity, linearity and resistance to dryness, and the analysis results are accurate and reliable.

6, improved quartz tube, reduce the production of carbon deposits, and eliminate pollution.

7. The instrument gas pipeline adopts imported passivity stainless steel pipeline to minimize the adsorption of the pipeline to the sample.

8. The sampler adopts wheeled injection, imported stepping motor, precise control, no noise when injecting.

 

Technical parameters

1. Sample type: liquid, solid and gas (need to be equipped with corresponding injector)

2, sample injection volume:

Solid sample: 1 ~ 5mg,

Liquid sample: 1 ~ 20uL,

3, measuring range: 0.5 ~ 10000mg / L ~ percentage

4, measurement accuracy:

0.2mg / L ≤ sample concentration <1.0mg / L ± 0.1mg / L

1.0mg / L ≤ sample concentration ≤ 100mg / L Cv ≤ 8%

100mg / L ≤ sample concentration ≤ 10000mg / L Cv ≤ 5%

5, temperature control range and accuracy: room temperature ~ 1100 ° C, ± 1 ° C

6, gas source requirements: high purity argon: purity 99.95% or more, high purity oxygen: purity 99.95% or more

7, power requirements: AC220V ± 22V, 50Hz ± 0.5Hz, 1500W

 

System configuration
Standard configuration:Printer+Computer+Analyzer+Liquid injector
Other optional parts:Solid injector,Gas injector
Based on Windows ( 7 , XP , Me, 98 ) English user interface to make the operation more convenient and fast.. You only need to click the mouse, and it can complete all of the parameter settings. The data collecting, processing, storage and printing controlled by computer.