SKYLINE INSTRUMENTS CO.,LTD info@skylineinstruments.com 86-769-23830463

SKYLINE INSTRUMENTS CO.,LTD Company Profile
products
Home > products > Oil Analysis Testing Equipment > SL-OA37 Cold Filtration Point Tester

SL-OA37 Cold Filtration Point Tester

Product Details

Place of Origin: CHINA

Brand Name: SKYLINE

Certification: CE , Calibration Certificate

Model Number: SL-OA37

Payment & Shipping Terms

Minimum Order Quantity: 1 Unit

Price: negotiated

Packaging Details: plywood case

Delivery Time: 15 working days

Payment Terms: L/C, D/P, T/T, Western Union, MoneyGram

Supply Ability: 10 unit / month

Get Best Price
Product Details
Highlight:

oil analysis lab equipment

,

oil analysis equipment

Keywords:
Oil Analysis Testing Equipment
Temperature range:
-60~50℃
Resolution:
1Pa
Pressure range:
0~200.0KPa
Temperature measuring element:
PT100 (temperature Sensor Of JUMO Company In Germany)
Ambient temperature:
5℃~40℃
Relative humidity:
≤85%
Power supply voltage:
AC 220V±10%50Hz±10%
Power consumption:
<900W
Dimensions:
600mm×450mm×4500mm
Instrument weight:
Host: 50kg, Suction Filter: 5kg
Keywords:
Oil Analysis Testing Equipment
Temperature range:
-60~50℃
Resolution:
1Pa
Pressure range:
0~200.0KPa
Temperature measuring element:
PT100 (temperature Sensor Of JUMO Company In Germany)
Ambient temperature:
5℃~40℃
Relative humidity:
≤85%
Power supply voltage:
AC 220V±10%50Hz±10%
Power consumption:
<900W
Dimensions:
600mm×450mm×4500mm
Instrument weight:
Host: 50kg, Suction Filter: 5kg
Product Description

SL-OA37 Cold Filtration Point Tester

SL-OA37 Cold Filtration Point Tester 0

Product description:

The cold filter point tester conforms to SH/T0248, and is suitable for measuring the cold filter point of distillate fuels, including diesel engine fuels containing flow improvers or other additives, and fuels used in civil heating devices.

 

Technical features:

1. Imported compressor refrigeration system ensures the required refrigeration depth

2. The built-in vacuum pump and electronic precision pressure balance system ensure that the suction filter pressure is automatically balanced at the set value

3. Automatically control the temperature difference between the cooling medium and the tested sample to ensure that the cooling rate is controlled and uniform

4. The internal precision microcomputer timing ensures the accuracy of the judgment results

 

Technical parameters:

Temperature range -60~50℃
Resolution 0.1°C
Pressure range 0~200.0KPa
Resolution 1Pa
Working cold tank single tank two baths, second bath isothermal
Temperature measuring element PT100 (temperature sensor of JUMO company in Germany)
Refrigeration method Compressor refrigeration (French DANFOS)
Timing 60.0s division 0.1s
Ambient temperature 5℃~40℃
Relative humidity≤85%
Power supply voltage AC 220V±10%50Hz±10%
Power consumption<900W
Dimensions
Host: 600mm×450mm×4500mm
Suction filter: 250mm×150mm×380mm
Instrument weight
Host: 50kg, Suction filter: 5kg